INORGANICCHEMICALSINDUSTRY ›› 2013, Vol. 45 ›› Issue (3): 48-.

Previous Articles     Next Articles

Simultaneous determination of arsenic and mercury in polyaluminum chloride by atomic fluorescence spectrometry method

Qin Xiaopeng1,Wu Kai1,Tao Futang2,Li Kai3   

  1. 1. Dongguan Huaqing Water Purification Technology Co.,Ltd.,Dongguan 523210,China;2.Taicang Xinxing Light
    Industrial Auxiliary Factory;3.Shenzhen Zhongrun Water Industry and Technology Development Co.,Ltd.
  • Online:2013-03-10 Published:2013-03-12

Abstract: Application of hydride generation-atomic fluorescence spectrometry method in the simultaneous determination of arsenic and mercury in polyaluminium chloride used for drinking water was studied.This method is of high sensitivity and good accuracy.Under optimal conditions,the potassium borohydride concentration,detection limit,precision,and recovery rate were investigated.Results showed:the detection limits of arsenic and mercury were 0.007 9 μg/L and 0.002 6 g/L, respectively:the recovery rates of arsenic and mercury were 93.75%~103.83% and 105.00%~108.75%,respectively.

Key words: hydride generation-atomic fluorescence spectrometry, polyaluminium chloride, arsenic, mercury