Detection of impurity elements in potassium dihydrogen phosphate with ICP-MS method

Ding Ling1,Liang Chiqiong2,Gong Chuangzhou1,Xia Junling1   

  1. 1.CNOOC Tianjin Chemical Research & Design Institute,Tianjin 300131,China;
    2.Foshan Supervision Testing Centre of Quality and Metrology
  • Online:2013-11-10 Published:2013-11-15

Abstract: 11 impurities analysis methods for determination of Be,Mg,Ni,Cu,Zn,As,Se,Cd,Sb,Tl,and Pb in potassium dihydrogen phosphate by inductively coupled plasma mass spectrometry(ICP-MS) was developed.After the sample was digested by HNO3,the above 11 impurities could be analyzed directly by ICP-MS.Li,Bi,In,and Rh were used to correct the matrix interference.Results showed the standard sample recovery rates of the 11 elements were in the range of 86.00%~118.77%.The relative standard deviation of the 11 elements were in the range of 3.79%~8.60%.The accuracy and precision of the methods both could meet the requirements.These methods are rapid,simple,and applicable for the analysis of various impurity elements in potassium dihydrogen phosphate.

Key words: inductively coupled plasma mass spectrometry, potassium dihydrogen phosphate, impurity elements